Step Parameters
Photometric spectrum scanning
96
Thermo Scientific SkanIt
The photometric spectrum scanning step includes the following parameters:
•
Scanning wavelengths - Select the starting and ending wavelengths
and the step size (scanning resolution) for the scanning measurement.
A step size is automatically suggested. The wavelength range is 200 to
1,000 nm.
•
Lag time [hh:mm:ss.t] - Define the time before the measurement is
executed.
•
Step duration [hh:mm:ss] - Define the total time for the step. The
software waits after the instrument has performed the action until the
time is up before proceeding with the following step.
•
Measurement time [ms] - The measurement or integration time defines
the light collection time used in the measurement. The wells are
measured with a selected measurement time. For more information, see
Measurement times" on page 113.
®
®