Average Rate - Thermo Scientific Varioskan User Manual

Multimode microplate reader software

Advertisement

Data Processing Parameters
Kinetic calculations

Average rate

184
Thermo Scientific SkanIt
Also known as normal rate. The average kinetic rate (slope of the absorbance
(Abs) vs. time curve) will be calculated by linear regression (linear least
squares method or LLS) using all the measurement readings within the
selected raw data and time range.
You can define the following settings:
Kinetic rate - Select to view the results in either seconds (s) or minutes
(min). The Kinetic rate is always calculated per second but if needed
the time is converted to minutes after the calculations.
Ignore first (n) readings - Ignores a defined number of points
(readings), counted from the first reading.
®
®
Software for Varioskan