Warning Messages - Thermo Scientific Varioskan User Manual

Multimode microplate reader software

Advertisement

Warning messages

Thermo Fisher Scientific
Code
Explanation
79
Bandwidth factors calibration
failed.
80
The measurement chamber door
is open.
81
Such an error during startup that
the execution of some functions
is not possible.
82
Invalid TRF measurement
command.
83
Varioskan LumiSens crosstalk
shield position is incorrect.
84
Varioskan LumiSens reference
light failure.
85
Dispensing head is not in the
correct position.
86
Wrong plate type.
88
Varioskan LumiSens base
installed, measurement module
not present.
Warning messages reported
Code
Explanation
100
Unable to comply with the
defined lag time.
101
Unable to comply with the
defined well interval.
102
Unable to comply with the
defined wavelength interval.
103
Unable to comply with the
defined kinetic interval.
104
Unable to comply with the
defined group interval.
105
The timer referenced in the
command is not running
(anymore). Your timing
requirement is not met.
Related to step duration.
Thermo Scientific SkanIt
Troubleshooting Guide
Suggested action
Contact service.
See that nothing is preventing the
measurement chamber door from
closing when the plate is driven in. Do
not open the door while the instrument
is measuring.
The measurement chamber door must
be closed during startup. If it is not,
contact service.
Contact service.
Clean the Varioskan LumiSens base.
Contact service if the error persists.
Contact service.
Move the dispensing head into the
correct position.
Use a suitable plate type. Dispensing
is only possible with up to 384-well
plates.
Install the Varioskan LumiSens
measurement module.
Suggested action
The minimum timing depends on the
combination of measurement
parameters, plate movement
parameters, dispensing parameters
and volume, number of wavelengths
and number of measured points.
®
®
Software for Varioskan
User Manual
Warning messages
273