Page 4
WEEE Conformity This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96 It is marked by the symbol opposite: Thermo Fisher Scientific has entered into agreements with recycling and disposal companies in all EU Member States for the recycling and disposal of this device.
Close lid ........................4-2 Rotor installation ......................4-2 Entering parameters ....................4-3 Saving programs .......................4-6 Centrifuging ......................4-7 Temperature adaptation during standstill ..............4-8 Short-term centrifuging ....................4-8 Removing the rotor ....................4-8 Aerosol-tight rotors ....................4-9 Switching the centrifuge off ..................4-9 Thermo Scientific SL 16 / 16R...
Page 6
Service of Thermo Fisher Scientific ................6-4 Chapter 7 Troubleshooting ........................7-1 Mechanical emergency lid release ................7-2 Troubleshooting by user ..................7-3 When to contact customer service ................7-6 Chemical Compatibility Chart ................A-1 Warranty ........................B-1 Contact Information ....................C-1 SL 16 / 16R Thermo Scientific...
Failure to follow the instructions and safety information in this instruction manual will result in the expiration of the sellers warranty. Scope of supply Article Number Quantity Check Centrifuge Thermo Scientific SL16 Power supply cable 50120986 Instruction manual 70009824 Anti-corrosion oil If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.
Accident Prevention Accident Prevention Prerequisite for the safe operation of the Thermo Scientific SL16 / 16R is a work environment in compliance with standards, directives and trade association safety regulations and proper instruction of the user: The safety regulations contain the following basic recommendations: •...
Page 9
CENTRIFUGE AND LEAVE THE AREA IMMEDIATELY. Normal use The centrifuge is to be used for separating materials of different density or particle size suspended in a liquid. Maximum sample density at maximum speed: 1,2 ---------- - Thermo Scientific SL 16 / 16R...
Page 10
WARNING means that injuries or material damage or contamination could occur. The symbol on the left refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe. SL 16 / 16R Thermo Scientific...
Introduction and Description Characteristics Characteristics The Thermo Scientific SL16 / 16R is an in-vitro diagnostics device according to the In-Vitro Diagnostics Directive 98/79/EC. Several rotors with commercial tubes can be used. The set speed is reached in seconds. The maintenance-free induction motor ensures quiet and low-vibration operation even at high speeds, and guarantees a very long lifetime.
Introduction and Description Technical Data Technical Data The technical data of the Thermo Scientific SL16 / 16R is listed in the following table. Table 1-1. Technical Data Thermo Scientific SL16 / 16R Feature Value Environmental conditions - use in interior spaces - altitudes of up to 2,000 m above sea level - max.
100V 50Hz IEC 61010-1 2 Edition IEC 61010-2-020 2 Edition IEC 61010-2-101 EN 61326-1 EN 61326-2-6 EN 55011A EN 61000-6-2 EN ISO 13485 SL 16 / 16R Thermo Scientific...
Automatic lid closing and locking starting from an initial hold position Mains Supply The following table contains an overview of the electrical connection data for the Thermo Scientific SL16 / 16R. This data is to be taken into consideration when selecting the mains connection socket.
Introduction and Description Rotor selection Rotor selection The Thermo Scientific SL16 / 16R is supplied without a rotor. Various rotors are available to choose from. TX-400 7500 3629 with round buckets 7500 3655 TX-200 7500 3658 with round buckets 7500 3659...
The set-up location must fulfil the following requirements: • A safety zone of at least 30 cm must be maintained around the centrifuge. People and hazardous substances must be kept out of the safety zone while centrifuging. SL 16 / 16R Thermo Scientific...
1. Turn off the power supply switch on the back (press "0"). 2. Plug the centrifuge into grounded electrical sockets only. 3. Check whether the cable complies with the safety standards of your country. Thermo Scientific SL 16 / 16R...
Before use Storage 4. Make sure that the voltage and frequency correspond to the figures on the rating plate. Figure 2-3. Rating plate of the Thermo Scientific SL16R 5. Establish the connection to the power supply with the connecting cable. Storage •...
Control panel The control panel contains the keys and displays of the centrifuge (only the power switch is located on the back of the device). All parameters can be selected and changed during operation. SL 16 / 16R Thermo Scientific...
Page 23
Control panel Control panel Thermo Scientific SL 16 / 16R...
Page 24
Control panel Control panel SL 16 / 16R Thermo Scientific...
Note The lid should audibly click into place. Rotor installation The approved rotors for the Thermo Scientific SL16 / 16R are listed in section “Rotor selection” page 1-6. Use only the rotors and accessories from this list in the centrifuge.
Page 27
4. Place the rotor lid onto the rotor. Be sure to check all sealings before starting any aeorsol-tight applications. See the information in the rotor instruction manual. 5. Close the centrifuge lid. Thermo Scientific SL 16 / 16R...
Entering parameters Entering parameters The Thermo Scientific SL16 offers you a total of 9 acceleration and 10 braking curves with which samples and gradients can be centrifuged. After the centrifuge is turned on, the last running profile selected is shown.
Page 29
1. Press the key below the TEMPERATURE display in order to open the temperature selection menu. The display shows "Set temp". 2. Enter the desired value by pressing the key repeatedly, until the desired value shows. Thermo Scientific SL 16 / 16R...
The display shows the following message: 2. Press the BUCKET key repeatedly until the bucket being used is displayed. Saving programs 1. Enter the program parameters. 2. Press any of the program store keys for 5 seconds. SL 16 / 16R Thermo Scientific...
The centrifuge will be deceler- ated at the designated rate. The message "END" will illuminate, and after pressing the OPEN key the lid will open and you can remove the centrifuged material. Thermo Scientific SL 16 / 16R...
Short-term centrifuging For short-term centrifuging, the Thermo Scientific SL16 / 16R has a "Pulse"- function. By holding down the PULSE key , spinning will start and continue until the key is let go.
Figure 4-5. AutoLock -lid for aerosol-tight rotors WARNING Mind the AutoLock -mandril inside the lid. Do not touch. Switching the centrifuge off 1. To turn off the centrifuge put the mains switch to "0". Thermo Scientific SL 16 / 16R...
Page 34
Operation Switching the centrifuge off 4-10 SL 16 / 16R Thermo Scientific...
To modify setting, use the keys. • Use the keys for the speed fields to navigate in the system menu. • Use the keys for the bucket field to navigate from one system menu item to the next. Thermo Scientific SL 16 / 16R...
Page 36
Start English Language English Deutsch Francaise End of run beep? Español Italiano Keypad beep? LCD powersave? Auto lid open? 0547V01 0547V01 Software ID: 4641V05 Software and NVRAM identification exam- ple values. Cycle count: 2368 SL 16 / 16R Thermo Scientific...
Drain off any melted ice from the rotor chamber. CAUTION Before using any cleaning or decontamination methods except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment. SL 16 / 16R Thermo Scientific...
9. Rinse the rotor and rotor lid thoroughly with water and then rub down. 10. Dispose of the disinfectant according to the applicable guidelines. 11. Treat aluminum rotors afterwards with an anti-corrosion oil (7000 9824). Apply a thin layer of oil using a soft cloth. Thermo Scientific SL 16 / 16R...
The service technicians check the following: • the electrical equipment • the suitability of the set-up site • the lid lock and the safety system • the rotor • the fixation of the rotor and the drive shaft SL 16 / 16R Thermo Scientific...
Page 41
Maintenance and care Service of Thermo Fisher Scientific Thermo Fisher Scientific offers inspection and service contracts for this work. Thermo Scientific SL 16 / 16R...
Page 42
Maintenance and care Service of Thermo Fisher Scientific SL 16 / 16R Thermo Scientific...
4. Push the cord back into the centrifuge and mount the plugs. Reconnect the centrifuge once the power has been restored. Switch on the centrifuge. Press the OPEN key to have the lid locks operative again. SL 16 / 16R Thermo Scientific...
The lid opens while the device is running. The run does not start or the centrifuge Close the lid and restart the centrifuge. runs down without being braked. If an error message appears again, inform Customer Service. Thermo Scientific SL 16 / 16R...
Page 46
Error during the self-test of the centrifuge program and the electronics. The run does not start or the centrifuge runs down without being braked. Restart the centrifuge. If an error message appears again, inform Customer Service. SL 16 / 16R Thermo Scientific...
Page 47
Error during the self-test of the centrifuge program and the electronics. The run does not start or the centrifuge runs down without being braked. Restart the centrifuge. If an error message appears again, inform Customer Service. Thermo Scientific SL 16 / 16R...
You enter the system menu. 2. Press the START key 3. Press and hold the ENTER key , until the following message is displayed: Software ID: XXXXXXXX 4. Communicate the software version to the service technician. SL 16 / 16R Thermo Scientific...
Thermo Fisher Scientific s prior written approval, or any use of replacement parts not supplied by Thermo Fisher Scientific, shall immediately void and cancel all warranties with respect to the affected Goods. Thermo Scientific SL 16 / 16R...
Page 58
+86 10 8419 3588 India +91 22 6716 2200 Japan +81 45 453 9220 Other Asian Countries +852 2885 4613 Latin America +1 866 984 3766 Other Countries +49 6184 90 6940 +33 2 2803 2180 SL 16 / 16R Thermo Scientific...
Page 59
Short-term centrifuging ............. 1-5 Features ........4-7 Starting centrifuge program ............. iii Foreword ......4-7 Stopping the centrifuging program ..............2-4 Storage ..........4-2 Switch on centrifuge ..........4-7 Imbalance indicator ..........2-2 Installation location ............1-3 Technical Data Thermo Scientific SL 16 / 16R...
Page 60
Index: ........4-8 Temperature adaptation ........2-2 Transporting the centrifuge SL 16 / 16R Thermo Scientific...