Thermo Scientific Varioskan User Manual page 107

Multimode microplate reader software

Advertisement

Thermo Fisher Scientific
Note If there are not any points selected in the multipoint
measurement, the protocol cannot be run.
Optics - Fluorescence can be measured from one position:
Top - Fluorescence is measured at the measurement position above
the plate.
Bottom - This option is available with the Varioskan Flash only.
Multiple measurements - Tick the check box to enter a multiple
wavelength pair and TRF parameter values. For more information, see
Defining multiple TRF measurements" on page 106.
TRF delay time [µs] - Define the waiting period (TRF delay) between
the excitation flash and the beginning of the emission light measurement
(see Figure 7-1).
TRF integration time [µs] - Define the length of the measurement
window, that is, the time the emitted signal is collected for each
measurement cycle (see Figure 7-1).
The sum of TRF delay time and TRF integration time can be maximum
2400 µs.
Thermo Scientific SkanIt
®
®
Software for Varioskan
User Manual
Step Parameters
TRF measurement
105