Download Print this page

Deleting Protocol Steps; Area Definition - Thermo Scientific Varioskan User Manual

Multimode microplate reader software

Advertisement

Deleting protocol steps

Area definition

Thermo Fisher Scientific
For more information on protocol steps, see Chapter 7: Step Parameters".
1.
Select (click) the protocol step you wish to delete.
If you delete a step that has child steps, such as the Area1 step in the
example above, the child steps (Photometric1 in the example) are deleted
too.
2.
Select Edit > Delete, or right-click and select Delete, or click the
button.
Caution Before selecting Edit > Delete, always check that you
have selected the item you wish to delete.
3.
Confirm the step deletion by clicking OK.
Note If the session has been executed, you can no longer edit it.
If you want to delete steps from the protocol, save the session with
a new name.
The Area Definition step is used to define the operation area for the steps
below it in the Steps step tree.
If there is no explicit area definition, steps apply by default to all the samples
in the layout. For steps that are not inside any area definition, the operation
area consists of all the wells that contain blanks, unknowns, controls and/or
calibrators.
You can select according to an area or a group.
Thermo Scientific SkanIt
®
®
Software for Varioskan
User Manual
Protocols
Area definition
85