To execute a session:
1.
Make sure that the instrument is connected correctly to the PC (see
Connecting the instrument to the PC through a COM port" on
page 37).
2.
Turn on the instrument (see the Thermo Scientific Varioskan User
Manual, Cat. no. 15018850).
3.
Start up the software (see Launching SkanIt Software" on page 54).
4.
Open the session (see Opening an existing session" on page 59). If
the session is already open, make sure that the session has been saved.
5.
Insert the plate into the plate tray (see the Thermo Scientific Varioskan
User Manual).
Caution Make sure you do not dispense into the instrument by
mistake. Ensure that:
A correct microplate has been inserted.
The microplate or tip priming vessel is not too full.
The recommended Check plate is ticked (Settings >
Options...).
6.
Prime the tubing if you have a dispensing step in the protocol (see
Priming dispensers" on page 75).
7.
Click
on the Execution action panel or select Execute > Session.
8.
Give a name to the run and click OK.
Thermo Scientific SkanIt
®
®
Software for Varioskan
User Manual
69