No. of dilutions - Enter the number of dilutions (not with calibrators
or controls).
No. of samples - Enter the number of samples.
No. of replicates - Enter the number of replicates.
Unit - Enter the unit for concentration or dilution volume.
You can let the software calculate the dilution/concentration values by
selecting Series, or enter individual values manually by selecting Multiple
values. See Example 8-2, Adding a concentration series" below.
If you select Series, the following parameters are available:
Initial value - Enter the initial concentration or dilution value.
Operators - Select the way in which the amount of reagent changes in
the series: the factor given in the Step By box is applied to the consecutive
concentration or dilution values according to the selection.
Step By - Enter the factor for the concentration/dilution series.
If you select Multiple Values, the following parameters are available:
Dilutions/Concentrations Enter each concentration or dilution value
into the small box and click Add. The value is added to the larger box.