Inspecting The Connecting Tubes And Leak Test Air Tube - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

3.4
Inspecting the connecting tubes and leak test air
tube
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Before using the equipment, always check that there is no irregularity regarding
the following points on the connecting tubes and leak test air tube.
 All tubes should be free of cracks, breaks, fissures,
 There should be no cracks in the lock levers of connecting
 There should be no bends or breaks in the pin of
 The tube should not be easy to disconnect once
Lock lever
Figure 3.2
If a tube has any irregularity, do not use it and replace with a new one.
Do not use the connecting tubes or leak test air tubes if they
have any irregularity. Doing so could prevent effective
reprocessing or damage the endoscope.
scratches, or stains.
tube connectors.
connecting tube connector.
connected.
Chapter 3 Inspection Before Use
Pin
43