Chapter 3 Inspection Before Use
3.6
Inspecting and replenishing alcohol
48
8.
Insert the white connector into the connector on the detergent tank until it
clicks.
9.
Turn the connector to correct the tube orientation as shown below. Confirm
that the tube is not bent.
Connector
Figure 3.9
10.
Close the detergent/alcohol drawer.
Check how much alcohol is in the alcohol tank and add more as required.
Check Required items
70% ethyl alcohol or isopropyl alcohol
Table 3.6
•
The alcohol used with the equipment must be 70% ethyl
alcohol or isopropyl alcohol. Using any other kind of alcohol
may result in malfunction of the equipment or the endoscope,
difficulty drying the endoscope, fire hazard, or a hazard due
to toxic vapor emitted from the alcohol.
•
Alcohol is flammable and should be handled with extra care.
•
Remove the alcohol in the alcohol tank and replace it with
new alcohol at least once a week. Otherwise, the alcohol in
the alcohol tank may degrade.
•
Before handling the alcohol, carefully read the precautions
for use carefully, and use the alcohol as instructed.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual