Care And Maintenance After Long-Term Storage; Setting Up The Equipment - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Chapter 7 Routine Maintenance

7.18 Care and maintenance after long-term storage

272
When using the equipment after it has been stored for more than 14 days
without being used, setting up the equipment, performing the reprocessing
program and performing the water supply piping disinfection are required.
Perform the following procedure.
 Setting up the equipment
1.
Connect the water supply hose. For detailed instruction, refer to Section 4.2,
"Connection of the water supply hose" in "Instructions-Installation Manual".
2.
Connect the power cord. For detailed instruction, refer to Section 4.4,
"Connection of the power cord" in "Instructions-Installation Manual".
3.
Confirm the power supply. For detailed instruction, refer to Section 4.5,
"Confirmation of the power supply" in "Instructions-Installation Manual".
4.
Install the detergent tank. For detailed instruction, refer to Section 3.13,
"Installation of the detergent tank" in "Instructions-Installation Manual".
5.
Add alcohol into the alcohol tank. For detailed instruction, refer to Section
4.10, "Addition of alcohol" in "Instructions-Installation Manual".
6.
Replace the water filter (MAJ-824). For detailed instruction, refer to
"Replacing the water filter" on page 195.
7.
Drain air in the water filter housing. For detailed instruction, refer to
"Draining air in the water filter housing" on page 199.
8.
Set up the disinfectant solution. For detailed instruction, refer to "Setting up
the disinfectant solution" on page 236 in this manual.
9.
Inspect and clean all parts before use. For detailed instruction, refer to
Chapter 3, "Inspection Before Use" in this manual.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual