Chapter 6 Other Functions
6.3

Setting the reprocessing programs

158
5.
When heating of the disinfectant solution is completed, the buzzer should
beep and the TIME/CODE display on the main control panel should show
[- -] indicating the end of the process.
Figure 6.4
Cleaning time can be set in reprocessing programs [2] and [3]. If the temperature
of the disinfectant solution in the reprocessing basin is below 20C (68F), it will
be heated to 20C (68F) before the disinfection process starts.
Cleaning time
Disinfecting time
Disinfectant
solution heating
Table 6.2
1 The OER-Pro is set for a 10-minute and 20C disinfection process for
ready-to-use disinfectant solution at the factory. Program changes to the
disinfection process will be released when Olympus-validated disinfectant
solutions with different disinfectant contact times and temperatures are
approved.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Program [1] setup
3 min
7 min (when using Olympus-validated concentrated disinfectant
10 min (when using Olympus-validated ready-to-use disinfectant
Program [2] and [3] setting
ranges
3 – 10 min. (in 1 min. steps)
1
solution)
1
solution)
1
(68F)
20C