Chapter 6 Other Functions
6.3
Setting the reprocessing programs
158
5.
When heating of the disinfectant solution is completed, the buzzer should
beep and the TIME/CODE display on the main control panel should show
[- -] indicating the end of the process.
Figure 6.4
Cleaning time can be set in reprocessing programs [2] and [3]. If the temperature
of the disinfectant solution in the reprocessing basin is below 20C (68F), it will
be heated to 20C (68F) before the disinfection process starts.
Cleaning time
Disinfecting time
Disinfectant
solution heating
Table 6.2
1 The OER-Pro is set for a 10-minute and 20C disinfection process for
ready-to-use disinfectant solution at the factory. Program changes to the
disinfection process will be released when Olympus-validated disinfectant
solutions with different disinfectant contact times and temperatures are
approved.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Program [1] setup
3 min
7 min (when using Olympus-validated concentrated disinfectant
10 min (when using Olympus-validated ready-to-use disinfectant
Program [2] and [3] setting
ranges
3 – 10 min. (in 1 min. steps)
1
solution)
1
solution)
1
(68F)
20C