Performing Leak Test Independently - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

3.
When the reprocessing basin is filled with water, the buzzer beeps three
times and the lid is unlocked.
4.
Make sure that the water supply has stopped and then step on the foot
pedal to open the lid.
5.
Check that there are not any bubbles produced continuously from the outer
surface of the endoscope and/or leak test air tube while angulating the
bending section of the endoscope. Check for leaks at least for 30 seconds.
6.
With the lid open and water in the basin, perform the procedure in Section
4.8, "Connecting tube installation" on page 108 and close the lid. The
reprocessing process will start.

Performing leak test independently

1.
Press the FUNC SEL button on the subcontrol panel to select "LEAK TEST".
Figure 4.28
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
When angulating the endoscope's bending section, be
careful not to let the bending section touch the reprocessing
basin or retaining rack as this could damage the endoscope.
After bending, straighten the bending section and place it
properly on the retaining rack.
When 10 minutes have passed since the leak test started,
the water will drain and the test will finish automatically. At
this time, the main control panel displays the error code
[E92].
Chapter 4 Reprocessing Operations
105