Olympus OER-Pro Operation Manual page 109

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

1.
Place the endoscope carefully, checking the following:
2.
Wipe the venting connector of the endoscope or the endoscope's water
resistant cap with a clean moistened with alcohol.
3.
Align the groove on the metal connector of the leak test air tube with the pin
on the venting connector of the endoscope or the endoscope's water
resistant cap, and turn the connector clock wise by 90 to connect it.
4.
If the leak test connector (black) in the reprocessing basin is wet, wipe the
entire connector with a clean cloth.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If an irregularity is found with the leak test air tube, replace it
with a new one and retry the leak test.
Before starting the leak test, always confirm that there is no
foreign material on the ventilation openings on the gas filter
cases. Blocking the ventilation not only hinders deodorization
but may also cause the device to malfunction.
Continuous production of air bubbles from a point on the
endoscope or the leak test air tube means that water may
penetrate at that point. If air bubbles are produced
continuously during leak test, discontinue leak test, withdraw
the endoscope or the leak test air tube from the reprocessing
basin and contact Olympus for endoscope service.
During leak test, the angulating section's covering may
expand. This is not a malfunction.
To discontinue the leak test while water is being supplied,
press the STOP/RESET button.
 The distal end of the insertion tube is straight.
 The distal end of the insertion tube is not on or beneath
another object.
 The leak test air tube is not twisted.
 The endoscope is not contacting the lid.
 The connecting tube is not connected.
Chapter 4 Reprocessing Operations
103