1.
Slowly open the water faucet.
2.
Make sure that the power switch is ON.
3.
Close the lid by pushing it until it clicks.
4.
Press the FUNC SEL button on the subcontrol panel to select "DRAIN
LCG".
Figure 7.64
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Do not push the disinfectant removal port with a finger while
the rubber cap is not attached. Otherwise, the disinfectant
solution may flow out.
To prevent the device and facilities near the equipment from
being damaged by leaked disinfectant solution, do not
remove the rubber cap from the disinfectant removal port
except when the disinfectant removal tube is connected.
If disinfectant solution leaks out of the disinfectant removal
port while the rubber cap is removed, immediately reattach
the rubber port and follow the procedure in Section 8.1,
"Troubleshooting guide" on page 274. If it does not stop
leaking, contact Olympus.