7.9

Cleaning the alcohol tank

ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
5.
Store the frequently used tubes in the accessory holder. Note that the
accessory holder cannot accommodate all of the tubes.
Figure 7.54
Do not hook tubes on the accessories holder in a disorderly
fashion as this could make it impossible to open or close the
front door and may also damage the accessories.
Do not tilt the alcohol or detergent tank while fluid is still
inside. Otherwise, the fluid may spill.
1.
Pull out the detergent/alcohol drawer.
2.
Disconnect the tube from the alcohol tank and take the tank out of the
detergent/alcohol drawer.
3.
Empty the tank.
4.
Rinse inside of the tank with running water. Repeat this rinse process
several times with fresh water to rinse it thoroughly.
5.
Using a cloth moistened with neutral detergent, clean the external surface,
rinse the inside and external surface of the tank in running water, and wipe it
with a clean cloth. To prevent bacterial growth, wipe the outside of the tank
with a cloth moistened with 70% ethyl alcohol or isopropyl alcohol.
6.
Drain out the water inside the tank, dry it thoroughly, put it back on the
detergent/alcohol drawer and attach the cap and tube to it.
Chapter 7 Routine Maintenance