Problem
Disinfectant solution is
judged to be ineffective with
the test strip.
Bacteria were detected as
a result of culture test of a
reprocessed endoscope.
Bacteria were detected as
a result of culture test of
rinsing water collected from
the device.
Endoscopes were not
precleaned before being
reprocessed.
Printed paper is not output
from the printer.
Abnormal noise from the
device.
The indicator lamp of the
communication status
indicator on the main
control panel goes
out/blinks.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Possible Causes
Expiration of the service life
of the disinfectant solution.
• Expiration of service life
of filters, degradation of
disinfectant solution, etc.
• Water supply piping is not
disinfected.
• Expiration of service life
of filters, degradation of
disinfectant solution, etc.
• Water supply piping is not
disinfected.
• Printer paper roll has run
out.
• Paper jam.
Internal problem with the
device.
The connection with
external device is
incomplete or a device
malfunctions.
Chapter 8 Troubleshooting and Repair
Remedial Actions
Replace the solution as described in Section 7.12,
"Replacing the disinfectant solution" on page 225.
Inspect the equipment as described in Chapter 3,
"Inspection Before Use", preclean the endoscope and
put it through the reprocessing process again from the
beginning. If bacteria are detected again in the next
culture test, contact Olympus.
Inspect the equipment as described in Chapter 3,
"Inspection Before Use". If bacteria are detected again in
the next culture test, contact Olympus.
Inspect the equipment as described in Chapter 3,
"Inspection Before Use". Then, preclean the
endoscopes and reprocess them again from the
beginning.
Take the remedial action by referring to Section 3.11,
"Inspecting the printer paper roll" on page 68.
Contact Olympus.
Check the power supply to the external devices and the
connection between the external devices and the
OER-Pro. If the problem cannot be resolved, contact
Olympus.