If error code [E94] is displayed during printing

1.
If the paper runs out while printing, the main control panel displays error
code [E94]. Press the STOP/RESET button on the main control panel to
clear the error code.
2.
Replace the printer paper roll.
3.
After replacing the printer paper roll, press the FUNC SEL button on the
subcontrol panel to select "PRINT TODAY".
4.
Press the FUNC START button on the subcontrol panel. The printer starts
printing the reprocessing results again from the beginning.
5.
If you want to print only the information after the point where printing was
stopped with error code [E94], do not clear the error code and simply
replace the printer paper roll.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If the printout is not satisfactory, printing can be restarted.
Press the FUNC SEL button on the subcontrol panel to select
"PRINT TODAY" and then press the FUNC START button.
The same information can be printed several times in the
same day.
For the printed information about "PRINT TODAY", refer to
"PRINT FORMAT (PRINT TODAY, PRINT ALL)" on
page 147.