Advertisement

Chapter 7 Routine Maintenance
256
6.
Detach the alcohol tank cap to which the tube is connected (Do not
disconnect the connector.) (see Figure 7.93).
Figure 7.93
7.
Remove alcohol from inside the line by performing the alcohol flush process
on the subcontrol panel. Check that the fluids are drained from the cap. Wait
until the error code [E93] is displayed and the process ends (see Figure
7.94).
Figure 7.94
8.
Prepare a beaker or similar container and pour about 80 ml of the
disinfectant solution into it.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual