Chapter 7 Routine Maintenance
204
When handling the disinfectant solution, wear personal
protective equipment to prevent any disinfectant from getting
on your skin or being inhaled. Avoid direct physical contact
and inhalation of vapors. If any disinfectant solution gets in
your eyes, immediately rinse with a large amount of fresh
water and then consult a medical specialist. Personal
protective equipment, such as eyewear, face mask,
moisture-resistant clothing, and chemical-resistant gloves
that fit properly and are long enough so that your skin is not
exposed. All personal protective equipment should be
inspected before use and replaced periodically before it is
damaged.
For details on replacing the disinfectant solution, see Section
7.12, "Replacing the disinfectant solution" on page 225.
After disinfection of the water supply piping, the quantity
and/or concentration of the disinfectant solution might be
reduced, which prevents additional reprocessing cycles from
being performed. Do not add fresh disinfectant solution to the
used disinfectant solution. Except for the case of performing
water line disinfection with fresh new disinfectant solution, do
not add fresh disinfectant solution to increase the volume of
the disinfectant solution in the OER-Pro. Drain the
disinfectant solution completely and replace with fresh
disinfectant solution after disinfection of the water supply
piping. It is recommended to perform disinfection of the water
supply piping immediately before routine replacement of the
disinfectant solution.
Disinfecting the water supply piping also disinfects the water
filter and other components in the piping.
1.
Check the concentration of the disinfectant solution as described in Section
3.10, "Checking the disinfectant solution concentration level" on page 59.
2.
Drain water from the water filter housing as described in "Draining water
from the water filter housing" in Section 7.2, "Replacing the water filter
(MAJ-824)" on page 192.
3.
Make sure that the water faucet is open.
4.
Step on the foot pedal to open the lid.