Olympus OER-Pro Operation Manual page 103

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

 When loading two 180 or older series endoscopes
As shown in Figure 4.19, tilt the endoscope connector of the second endoscope
and put it in place while slightly lifting the endoscope connector of the first
endoscope to the left (direction of arrow 1). Place the scope ID tag to the position
shown in Figure 4.19.
Endoscope connector of
the first endoscope
1
Figure 4.19
Figure 4.20
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
For setup of "OSF-V60","two channel endoscope", and
"ultrasonic endoscope", refer to the "OER-Pro Quick
Reference Guide".
When loading only one conventional endoscope or two
conventional endoscopes that are 180 series or older, it is
possible to place endoscope connector without tilting. In this
case, the endoscope connector of the second endoscope
should be placed on top of that of the first endoscope as
shown in Figure 4.20. Then, place the scope ID tag of the
second endoscope to the position shown in Figure 4.20.
Chapter 4 Reprocessing Operations
Put the endoscope
connector of the second
endoscope in this area
Tilt the endoscope
connector of the second
endoscope
Scope ID tag
Place the universal cord on
the inside of the hook
Place the endoscope connector of
the second endoscope on top of the
endoscope connector of the first
endoscope
Scope ID tag
Place the universal cord on
the inside of the hook
97