Olympus OER-Pro Operation Manual page 179

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

1.
Make sure that the water faucet is open.
2.
Make sure that the lid is closed.
3.
Check the ALCOHOL indicator on the equipment's detergent/alcohol drawer
to confirm that the alcohol level is above "MIN.". If the amount of alcohol is
below the "MIN." line, add more alcohol as described in Section 3.6,
"Inspecting and replenishing alcohol" on page 48.
 Performing alcohol flushing independently
1.
Press the FUNC SEL button on the subcontrol panel to select "ALC
FLUSH".
2.
Press the FUNC START button on the subcontrol panel. The TIME/CODE
display of the main control panel shows [03], which indicates 3 minutes. The
time displayed on the main control panel counts down every minute.
3.
When the alcohol flushing process finishes, the buzzer beeps and the
TIME/CODE display shows [- -].
Figure 6.23
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Alcohol flushing includes the water feed and drain
operations. These operations are intended to drain alcohol
while diluting it.
Chapter 6 Other Functions
173