Chapter 3 Inspection Before Use
3.5

Inspecting and replacing the detergent tank

Inspection of the amount of detergent

44
Check Required items
Olympus-validated detergent
Table 3.5
Contact Olympus to obtain the list of Olympus-validated
detergents.
Check the DETERGENT indicator on the detergent/alcohol drawer of the device
to confirm that the amount of the detergent in the detergent tank. If the level
reaches "MIN", replacement of the detergent tank is required within a few
reprocessing operations.
Replacement of the detergent tank is required only when the error code [E95] is
displayed on the main control panel (and the detergent replacement indicator
lights up) after the process is started. In that case, replace detergent tank as
described in "Replacing the detergent tank" on page 45.
Figure 3.3
The detergent tank can hold about 2 L of detergent (which
can be used for about 30 reprocessing operations).
If reprocessing is initiated without detergent, the error code
[E95] is displayed and the reprocessing is stopped. The
Detergent replacement indicator on the main panel blinks.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual