Alcohol flushing

172
This process automates the flushing of the endoscope channels with alcohol
followed by air to help dry the channels.
When using the disinfectant solution and alcohol, Olympus
recommends the use of gas filters and running this
equipment in well-ventilated areas.
 Wear a face mask, gloves, and protective clothes to
 Wear goggles for eye protection.
Refer to the following association's guidelines related to
ventilation:
Do not handle the equipment if the operator shows any
allergic symptoms even while wearing protective gear.
When the alcohol flushing process is stopped due to an
equipment error, do not use the endoscope and start the
alcohol flushing process again from the beginning.
Otherwise, alcohol may remain in the endoscope channel
and pose a risk to patient safety.
Do not perform alcohol flushing without connecting the
connecting tubes. Otherwise, excessive pressure on the
pipes in the equipment may damage it.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
minimize aspiration and skin contact.
SGNA
(Society of Gastroenterology Nurses and
Associates)
ASGE
(American Society of Gastroenterological
Endoscopy)
APIC
(Association for Professionals of Infection
Control and Epidemiology)
AORN
(Association of Preoperative Registered
Nurses)
ASTM
(American Society for Testing and Materials)
OSHA
(Occupational Safety and Health
Administration)
ACGIH
(American Conference of Governmental
Industrial Hygienists)
NIOSH
(National Institute for Occupational Safety
and Health)
AIA
(American Institute of Architects)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents