ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
25.
Disconnect the water supply piping disinfection hose and close the lid.
Water supply piping disinfection hose connector
Washing case mount
Figure 7.36
26.
Rinse the filter tube and water supply piping disinfection hose thoroughly in
running water, dry them thoroughly and store in a clean place.
Perform microbiological sampling of the OER-Pro rinse water quality as required
by your hospital's policy. Also Olympus recommends to perform microbiological
sampling of the rinse water right after performing the water supply piping
disinfection if the device has not been used for more than 14 days. Sampling
from the inner piping of the equipment should be conducted by using the
following methods.
Check
Sampling tube (gray)
Container with 2 L or large capacity (wide-mouthed container such as a vat)
Sterilized bottle
Table 7.8
Chapter 7 Routine Maintenance
Air/water/instrument
channel connector
Required items
211