Advertisement

5.
Hold the disinfectant collection hose so that it does not move, then press the
FUNC SEL button on the subcontrol panel to select "DRAIN LCG".
Figure 7.58
6.
Press the FUNC START button on the subcontrol panel. The buzzer repeats
short beeps to indicate that the disinfectant solution drain process is
underway, the TIME/CODE display on the main control panel shows a
spinning [
7.
When the container (large) becomes nearly full, press the STOP/RESET
button on the main control panel to interrupt the disinfectant solution
collection.
8.
Prepare a new container (large) and put the other end of the disinfectant
collection hose in it.
9.
Hold the disinfectant collection hose so that it does not move, then press the
FUNC SEL button on the subcontrol panel to select "DRAIN LCG", and then
press the FUNC START button to restart draining of the disinfectant
solution.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
] as shown below, and the disinfectant solution begins draining.
Chapter 7 Routine Maintenance
229