Replacing The Gas Filter (Maj-822) - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Chapter 7 Routine Maintenance
7.1

Replacing the gas filter (MAJ-822)

188
Check Work to be performed as required
7.10 Checking cassette cutters (when using concentrated disinfectant solution)
7.11 Cleaning the disinfectant bottle drawer (when using concentrated
disinfectant solution)
7.12 Replacing the disinfectant solution
7.13 Cleaning the mesh filter in the water supply hose connector
7.14 Replacing the fuse
7.15 Disinfecting the internal detergent/alcohol lines
7.16 Preparing the reprocessor for long-term storage
7.17 Installing the printer paper roll
7.18 Care and maintenance after long-term storage
Table 7.3
Replace the gas filter every month or whenever the odor of the disinfectant
solution seems to have increased, whichever comes first.
Check Required items
Gas filter (MAJ-822) ( 2 pieces)
Indelible pen or other writing implement
Table 7.4
The disinfectant vapor generated by the device has been
proven safe in in-house testing. Nevertheless, as individual
reactions to the disinfectant may vary, Olympus recommend
the use of gas filters and enhanced protection by observing
the ventilation conditions given in "Ensuring the safety of
reprocessing personnel" on page 11. The use of the gas filter
does not replace the need for proper room ventilation.
Do not block the ventilation openings on the gas filter cases
with the replacement date indication sticker or any other
foreign material. Blocking the ventilation not only hinders
deodorization but may also cause the device to malfunction.
If the odor of the disinfectant solution seems to have
increased after replacement of gas filters, contact Olympus.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual