Chapter 3 Inspection Before Use
3.3

Inspecting the connectors

42
Check the following for each connector.
 The connector should be fixed firmly.
 The O-rings should be free of abnormalities such as
If any irregularity is found, do not use the equipment and contact Olympus.
Check Connectors in the reprocessing basin
Air/water/instrument channel connectors
Auxiliary water/elevator wire channel connectors
Leak test connector
Water supply piping disinfecting connector
Table 3.3
Check Other connectors
Connector above water filter housing
Connector below water filter housing
Disinfectant removal port
Tube connector on detergent tank
Tube connector on alcohol tank
Table 3.4
Do not use the equipment if any connector seems to be
damaged or defective. Using the equipment when an
irregularity has been detected may interfere with
reprocessing. Furthermore, fluid leakage may damage
peripheral devices or facilities near the equipment.
Connect each connector firmly by pushing until the connector
clicks into place. After connection, pull the connector gently
to confirm that it cannot be disconnected easily.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
cracks, tears, or dents.