8.
When water starts to flow continuously from the filter tube, disconnect the
tube by pushing its lock lever. Water flow should stop when the filter tube is
disconnected.
Figure 7.22
9.
Make sure that no water leaks from the water filter housing. If a water leak is
observed, immediately press the STOP/RESET button to stop water supply
and reinstall the water filter (restart procedures beginning with "Draining
water from the water filter housing" on page 193).
10.
Press the STOP/RESET button to drain water from the reprocessing basin.
11.
Close the front door.
12.
Rinse the filter tube with running water, dry it completely and store in a clean
place.
13.
Go to Section 7.3, "Disinfecting the water supply piping" on page 202 and
follow the instructions.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Chapter 7 Routine Maintenance
201