Advertisement

Chapter 7 Routine Maintenance
232
15.
Hold the lock levers and slowly disconnect the tube (see Figure 7.63).
Figure 7.63
16.
Rinse the disinfectant removal tube thoroughly under running water, dry it
completely, and store it in a clean place.
17.
Wipe the disinfectant removal port with the clean cloth and attach the rubber
cap to it.
18.
Close the front door.
The front door cannot be closed unless the rubber cap is
attached.
 Draining through the drain hose
Check Required items
Disinfectant removal tube
Container (small) with a capacity of about 200 ml, such as a beaker
Clean cloth
Table 7.14
Be sure to close the lid before proceeding. If the lid is not
closed, the disinfectant solution may spatter out of the
reprocessing basin.
Remove all endoscopes, valves, and connecting tubes from
the reprocessing basin before draining the disinfectant
solution through the drain hose. Otherwise, the disinfectant
solution cannot be drained properly and the endoscopes and
valves may be unable to be rinsed sufficiently.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Lock levers