Chapter 3 Inspection Before Use
3.8

Inspecting for disinfectant solution odor

54
4.
If any foreign object is found to be clogging the filter, clean the mesh filter in
running water using a soft brush.
Figure 3.18
5.
Attach the mesh filters in their original positions.
Check that the disinfectant solution is not producing an abnormal odor.
Before handling the disinfectant solution, carefully read the
precautions attached to the disinfectant solution bottle and
use as instructed.
1.
Activate the room's ventilation system.
2.
Check that there is no abnormal disinfectant solution odor coming from the
equipment or its surroundings.
If the odor increases after replacement of the gas filters,
contact Olympus.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual