Chapter 3 Inspection Before Use
Gas filter case mount
66
5.
Attach the gas filter case to the dome part of the lid.
Figure 3.32
6.
Rinse the test strip forceps thoroughly in running water, dry it thoroughly and
store in a clean place.
 Restarting the reprocessing after stopping due to
disinfectant solution failure
If the concentration of the disinfectant solution was checked during the
reprocessing and the disinfectant solution failed to meet the minimum
recommended concentration, stop the reprocessing and replace the disinfectant
solution. Perform the reprocessing cycle again from the beginning.
Always wear appropriate personal protective equipment,
such as eyewear, face mask, moisture-resistant clothing, and
chemical-resistant gloves that fit properly and are long
enough so that your skin is not exposed. Otherwise, blood
and mucus attached to the endoscopes could cause an
infection.
Replace the disinfectant solution when it fails to meet the
minimum recommended concentration. Otherwise, the
reprocessing may be insufficient.
Once the disinfectant solution failed to meet the minimum
recommended concentration, immediately stop the
reprocessing process. Otherwise, the reprocessing logs will
be incorrect.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual