Olympus OER-Pro Operation Manual page 67

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

1.
Push [PUSH] on the front door to open the front door. Remove the rubber
cap from the disinfectant removal port.
Figure 3.26
2.
Push the drain connector into the disinfectant removal port until it clicks.
3.
Place the prepared container below the drain connector, push the
connector's valve, and collect the drained disinfectant solution until it is
deep enough to immerse the reaction area of a test strip.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
To obtain disinfectant test strips, contact Olympus.
Disinfectant
removal port
Rubber cap
When connecting the drain connector to the disinfectant
removal port, do not push on the connector's valve.
Otherwise, disinfectant solution will leak out of it.
Chapter 3 Inspection Before Use
61