Olympus OER-Pro Operation Manual page 71

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

3.
Remove the gas filter case from the dome part of the lid.
Figure 3.30
4.
Insert the test strip through the gas filter case mount and immerse it in the
disinfectant solution for the prescribed period of time. In order to properly
immerse the test strip in the disinfectant solution, insert the test strip when a
jet of disinfectant solution from the water/circulation nozzle is not hitting the
dome of the lid. If the concentration is below the effective level, press the
STOP/RESET button immediately to stop the reprocessing cycle. Perform
rinsing, removal of the endoscopes, replacement of the disinfectant solution,
and then start the reprocessing cycle again form the beginning as described
in "Restarting the reprocessing after stopping due to disinfectant solution
failure" on page 66.
Figure 3.31
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Chapter 3 Inspection Before Use
65