Immunity test
IEC 61326 test level
Contact:2, 4 kV
Electrostatic
discharge (ESD)
Air:2, 4, 8 kV
IEC 61000-4-2
Electrical fast
2 kV
transient/burst
for power supply lines
IEC 61000-4-4
1 kV
for input/output lines
Surge
Differential mode:
IEC 61000-4-5
1 kV
Common mode:
2 kV
Voltage dips,
< 5% U
(>95% dip in U
short interruptions,
and voltage
for 0.5, 1 cycle
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
30 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
 Electromagnetic immunity compliance information and
recommended electromagnetic environments
Compliance level
Same as left
Same as left
Same as left
Same as left
T
)
T
Same as left
U
T
test level.
Floors should be made of wood, concrete, or
ceramic tile that hardly produces static. If floors
are covered with synthetic material that tends to
produce static, the relative humidity should be
at least 30%.
Mains power quality should be that of a typical
commercial (original condition feeding the
facilities) or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of this equipment required continued operation
during power mains interruptions, it is
recommended that this equipment be powered
from an uninterruptible power supply or a
battery.
It is recommended to use this equipment by
maintaining sufficient distance from any
equipment that operates with high current.
is the a.c. mains power supply prior to application of the
Appendix
Guidance