6.7

Air purge

ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Air purge is performed to drain the remaining cleaning fluid and disinfectant
solution from the basin after an irregularity occurs, or if the process is stopped
before it completes. Air purge also eliminates residual fluid from the endoscope
channels.
1.
Close the lid.
2.
Press the FUNC SEL button on the subcontrol panel to select "AIR
PURGE".
Figure 6.20
3.
Press the FUNC START button on the subcontrol panel to start air purge.
The TIME/CODE display on the main control panel shows [10], which
indicates 10 minutes. The elapsed time displayed on the main control panel
counts down every minute.
4.
The buzzer beeps and air purge is completed in 10 minutes. The main
control panel displays [- -] to indicate the end of air purge.
To stop the air purge process midway, press the
STOP/RESET button on the main control panel.