Olympus OER-Pro Operation Manual page 163

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

1.
Close the lid.
2.
Press the FUNC SEL button on the subcontrol panel to select "HEAT LCG".
3.
Press the FUNC START button on the subcontrol panel. The heat
disinfectant indicator will light to indicate that the operation is in progress
and the TIME/CODE display on the main control panel shows a spinning
[
] mark as shown below. Disinfectant solution is then poured into the
reprocessing basin and the heating process starts.
Figure 6.3
4.
When the disinfectant solution temperature reaches the specified level, the
disinfectant solution is automatically collected and the reprocessing basin is
rinsed.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
The disinfectant solution temperature takes about 4 minutes
for each 1C increase.
The disinfectant solution is heated to slightly above 20C
(68F) to compensate for the decrease in temperature inside
the equipment.
Chapter 6 Other Functions
157