 Setting workflow
Determine the disinfection operation count and elapsed
day count until the disinfectant solution expires.
When the disinfectant check shows that the
disinfectant is no longer effective, confirm the
disinfection operation count and elapsed day count.
Register the disinfectant solution expiry conditions.
Enter the disinfectant operation count and
elapsed day count identified above.
The disinfectant counter lamp blinks.
The lamp lights when either the set expiry
condition (disinfectant operation count or
elapsed day count) is reached.
Update the set values as required.
Modify the set values according to changes
in the operating environment, etc. (The
equipment does not automatically determine
the ageing of the disinfectant solution.)
Figure 6.14
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Chapter 6 Other Functions
Be sure to check the
disinfectant
concentration using
the Test Strip every
time you use the
reprocessor.
165