Chapter 7 Routine Maintenance
7.7

Cleaning the detergent/alcohol drawer

220
5.
Wipe the moisture around the float switch using a clean cloth.
6.
Attach the float switch cover to the original position above the float switch.
Figure 7.51
Be careful not to injure your hand by hitting the
detergent/alcohol drawer.
1.
Pull out the detergent/alcohol drawer.
2.
Disconnect the tubes from the detergent tank and alcohol tank, and take the
tanks out of the drawer.
3.
Take the detergent/alcohol inner tray out of the detergent/alcohol drawer.
4.
Clean the detergent/alcohol inner tray in fresh running water.
5.
After cleaning, dry it thoroughly with a clean cloth.
6.
Place the detergent/alcohol inner tray back on the detergent/alcohol drawer.
7.
Place the detergent tank and alcohol tank on the detergent/alcohol drawer
and connect the tubes to the original positions on the tanks.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual