PRINT FORMAT (PRINT TODAY, PRINT ALL)

Figure 4.59
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
If the "Error Result Printout" is set to "exclude"
PRINT TODAY
The reprocessing results recorded that day are
printed in chronological order.
PRINT ALL
All the reprocessing results stored in the equipment
are printed in chronological order.
If the "Error Result Printout" is set to "include"
PRINT TODAY
The reprocessing results and the error results
recorded that day are printed in chronological order.
PRINT ALL
All the reprocessing results and the error results
stored in the equipment are printed in chronological
order.
End of printing
If the RFID function has not been enabled at the factory, the
columns for the serial numbers of endoscopes, model names
of endoscopes, user number, and user name will be left
blank.
If the date and/or time setting has been changed in the past
than the date and/or time of the latest record stored in the
equipment, the reprocessing records of the past may not be
printed or not be printed in chronological order. Before
changing the date and/or time setting, print records of the
past stored in equipment by print function.