Chapter 4 Reprocessing Operations
4.1 Power activation and opening the water faucet
4.2 Inspection before use
4.3 Endoscope precleaning
and manual cleaning
4.4 Scope ID detection (When using the RFID function)
4.5 User ID detection (When using the RFID function)
4.6 Endoscope setup
4.7 Leak test
4.8 Connecting tube installation
4.9 Reprocessing
4.10 Removing the endoscopes
4.11 Printing of the reprocessing records
Chapter 5 End-of-Day Checks
74
(Follow the cleaning, disinfection, and sterilization procedures
described in the endoscopes' instruction manual or use modified
precleaning described in section 4.3. When reprocessing Gyrus
ACMI endoscopes with the OER-Pro, the Gyrus ACMI endoscopes
should be manually cleaned according to the procedure
described in the instruction manual for MAJ-1904.)
Connection of leak test air tube
Required leak test time:
Approximately 5 minutes
Program (can be selected from three reprocessing programs)
Required time: Approximately 26 minutes (concentrated disinfectant solution) or
29 minutes (ready-to-use disinfectant solution) at the shortest
Water supply Ultrasonic cleaning Detergent cleaning Rinsing Air purge 
Disinfectant solution heating Disinfection using disinfectant solution  Rinsing 
Air purge Alcohol flushing
Disconnection of leak test air tube and connecting
tubes, removal of endoscopes
Figure 4.1
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Inspection at the start of the work day
Inspection before every operation
• Inspection of pump function and piping
• Check of disinfectant solution concentration
Required items • Test strip
1
1
1
Required items
• Leak test air tube (MAJ-821)
• 70% ethyl alcohol or isopropyl alcohol
• Clean cloth
Required items
• Connecting tubes
(See "List of Compatible
Endoscopes/Connecting Tubes <OER-Pro>".)
Required items
• Sterilized cloth such as gauze
: Mandatory operations
: Optional operations
: Automatically performed operations
• Drain connector
• 200 ml cup-shape
container
• Clean cloth
This function is disabled when
the equipment is shipped.
If you want to use this function,
please contact Olympus.