3.12 Inspecting the communication status
(for connecting with external devices)
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Inspect the communication status with external devices. This procedure is
applied only when the communication is activated.
If the communication between the OER-Pro and external
device is interrupted, the reprocessing data will be stored in
the OER-Pro without being transmitted to external devices. If
the communication is kept interrupted, the older data in the
OER-Pro may be lost without transmitted to external devices.
Immediately resolve the problem when the lamp blinks/goes
out to prevent the loss of reprocessing data. If you want to
continue the reprocessing program, although the lamp is
blinking, print the reprocessing data (including reprocessing
results and error results). Otherwise, the loss of reprocessing
data may occur.
Always use Olympus designated external devices for
communication with the OER-Pro. Olympus will not support
the use of non-designated devices. For more details on
external devices available, contact Olympus.
The OER-Pro can store up to 60 reprocessing data (60
reprocessing results and 60 error results). The
communication status indicator starts to blink in order to
notify that the communication between the OER-Pro and the
external devices has been interrupted and the OER-Pro is
nearing its maximum capacity for data storage.