Olympus OER-Pro Operation Manual page 221

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

11.
Press the FUNC START button on the subcontrol panel (see Figure 7.43).
Figure 7.43
12.
Make sure that water comes out from the end of the sampling tube and the
sampling tube is filled with water (see Figure 7.44).
Figure 7.44
13.
Hold the end of sampling tube in a sterile collection container. Collect an
amount of water required for the microbiological examination.
14.
After collection, press the STOP/RESET button twice, and confirm the
TIME/CODE display on the main control panel shows the reprocessing time
(If the TIME/CODE display on the main panel shows [E00], press the
STOP/RESET button again).
15.
Step on the foot pedal to open the lid.
16.
Place the sampling tube on the retaining rack (see Figure 7.37).
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Be sure to press the STOP/RESET button twice. Otherwise,
the water may spray out of the sampling tube and leak from
the container during the automatic processing.
Chapter 7 Routine Maintenance
215