Chapter 7 Routine Maintenance
264
23.
Attach the connector jigs to the connectors on the reprocessing basin.
Figure 7.106
24.
Close the lid.
25.
Perform the operation described in Section 6.7, "Air purge" on page 169.
26.
Close the water faucet.
27.
Press the FUNC SEL button on the subcontrol panel to select "LEAK TEST",
and press the FUNC START button.
Figure 7.107
28.
In about 10 seconds, press the STOP/RESET button to stop the leak test.
29.
Drain water from the water filter housing as described in "Draining water
from the water filter housing" on page 193.
30.
Step on the foot pedal to open the lid, disconnect the connecting tubes, dry
them thoroughly and store them in a clean place.
31.
Press the power switch to set it to OFF.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual