Chapter 7 Routine Maintenance
260
5.
Connect the provided syringe and tube.
Figure 7.99
6.
Connect the tube attached to the syringe to the detergent nozzle inside the
reprocessing basin.
7.
To remove detergent from inside the line, connect the tube attached to the
syringe to the detergent nozzle inside the reprocessing basin. Suction the
detergent with the syringe until no more detergent comes out of the line (see
Figure 7.100).
Figure 7.100
8.
Prepare a beaker or similar container and pour more than 50 ml of the
sterile water into it.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual