Inspecting The Lid And Lid Packing; If Water Leaks From The Drain Hose Connector - Olympus OER-Pro Operation Manual

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

3.2

Inspecting the lid and lid packing

ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
 If water leaks from the drain hose connector
1.
Close the water faucet.
2.
Check the drain hose installation by referring to Section 4.3, "Connection of
the drain hose" in "Instructions-Installation Manual".
Before using the equipment, always check that there is no irregularity regarding
the following points on the lid and the lid packing. If there is any irregularity,
cleaning fluid or disinfectant solution may leak out.
The lid is not cracked, broken, or otherwise damaged.
The packing is not cracked, torn, or otherwise damaged.
The packing is not separated or detached from the lid.
The lid can be properly opened and closed without rattle.
Figure 3.1
If any irregularity is found, do not use the equipment and contact Olympus.
Do not use the equipment if the lid or the lid packing seems
to be damaged or defective. Using the equipment when an
irregularity has been detected may interfere with
reprocessing. Furthermore, fluid leakage may damage
peripheral devices or facilities near the equipment. If any
irregularity is observed with the lid or the lid packing, please
contact Olympus.
Chapter 3 Inspection Before Use
Packing
41