Olympus OER-Pro Operation Manual page 244

Endoscope reprocessor
Hide thumbs Also See for OER-Pro:
Table of Contents

Advertisement

Chapter 7 Routine Maintenance
238
Follow the disinfectant manufacturer's instructions for any
preparation or activation required prior to loading into the
OER-Pro.
 When using Olympus-validated concentrated disinfectant
solution
Check Required items
Acecide-C high level disinfectant solution
Table 7.15
Do not put your hand into the disinfectant bottle drawer or
near the cassette bottle cutters. Irritation of skin due to
contact with concentrated disinfectant solution, injury by
touching a projection, or malfunction of this equipment may
result.
If the disinfectant solution in the cassette bottles has not
completely drained after preparation of disinfectant solution,
do not use the equipment and contact Olympus.
Inappropriate preparation of disinfectant solution will prevent
proper endoscope reprocessing.
Do not touch inside the caps of cassette bottles. Do not push
or apply strong pressure to the bottle. Otherwise, disinfectant
solution may leak from the bottle.
To avoid malfunction, do not attempt to pull out the
disinfectant bottle drawer while it is locked.
To obtain Acecide-C high level disinfectant solution, contact
Olympus.
1.
Make sure that the water faucet is open.
2.
Make sure that the power is ON.
3.
Close the lid.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual