Chapter 7 Routine Maintenance
7.13 Cleaning the mesh filter in the water supply hose
connector
248
If the buzzer beeps and the TIME/CODE display shows [- -],
stop adding the disinfectant solution. Otherwise, the
disinfectant solution may spill from the equipment or be
drained from the equipment.
8.
When the amount added is adequate, the buzzer beeps for three seconds.
Stop adding the solution. At this time, the TIME/CODE display on the main
panel will show [--] indicating the end of the process (see Figure 7.83).
Figure 7.83
9.
Make sure that the water faucet is open.
10.
Perform the rinsing process following the procedure in Section 6.8, "Rinsing"
on page 170. Any residual disinfectant solution in the reprocessing basin is
removed by rinsing.
When the equipment stops with error code [E01], the water filter should be
replaced first. However, if the equipment stops again with error code [E01], clean
the mesh filter as described below.
1.
Close the water faucet. Press the FUNC SEL button on the subcontrol panel
to select "LEAK TEST", and press the FUNC START button to relieve the
incoming water pressure.
2.
After the pressure has been relieved (approximately 10 seconds), press the
STOP/RESET button to stop the leak test.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual