9.
Put the detached cap in the container filled with the disinfectant solution
(see Figure 7.95).
Figure 7.95
10.
Fill the alcohol supply line with the disinfectant solution by performing the
alcohol flush process on the subcontrol panel. Check that air does not get
inside the alcohol supply line. Also check that the disinfectant solution in the
container is decreasing as the process progresses. The process ends
automatically in three minutes.
Figure 7.96
11.
Maintain the disinfectant solution in the supply line for the amount of time
required for disinfection. Refer to the documentation supplied with the
disinfectant solution for specifics regarding disinfection.
12.
When the disinfection time ends, take the cap out of the container filled with
the disinfectant solution.
13.
Remove disinfectant solution from inside the line by performing the alcohol
flush process on the subcontrol panel. The process ends automatically with
the E93 displayed on the main control panel.
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Chapter 7 Routine Maintenance