Chapter 7 Routine Maintenance
194
8.
Press the FUNC START button on the subcontrol panel.
Figure 7.9
Connector below water filter housing
Figure 7.10
9.
When the water flow stops, press the STOP/RESET button to end the air
purge process, and disconnect the two filter tubes by pushing the lock levers
on their connectors.
Figure 7.11
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Connector above water filter housing
Lock lever